首页   按字顺浏览 期刊浏览 卷期浏览 Low-coherence lock-in interferometer for ophthalmology
Low-coherence lock-in interferometer for ophthalmology

 

作者: A. C. Oliveira,   F. M. M. Yasuoka,   J. B. Santos,   L. A. V. Carvalho,   J. C. Castro,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1998)
卷期: Volume 69, issue 4  

页码: 1877-1881

 

ISSN:0034-6748

 

年代: 1998

 

DOI:10.1063/1.1148867

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The measurement of short distances by interferometry using low coherence light is a potential option in the study of interfaces between intraocular elements. In this article we describe an experimental device to measure the distance between such interfacesin vitro. The procedure utilizes a Michelson interferometer to generate the probe beams together with a silicate photodetector system and a lock-in synchronous amplifier. The cornea and crystalline lens thickness of a rabbit eyeball are evaluated, as well as the human cornea in preservative solution. These tests indicate the method to be efficient, easy-to-handle, of high resolution and to present a signal-to-noise ratio of up to 12. We consider the possibility of its use in obtaining information about the degeneration of intraocular structures during the preservation process. ©1998 American Institute of Physics.

 

点击下载:  PDF (805KB)



返 回