Low-coherence lock-in interferometer for ophthalmology
作者:
A. C. Oliveira,
F. M. M. Yasuoka,
J. B. Santos,
L. A. V. Carvalho,
J. C. Castro,
期刊:
Review of Scientific Instruments
(AIP Available online 1998)
卷期:
Volume 69,
issue 4
页码: 1877-1881
ISSN:0034-6748
年代: 1998
DOI:10.1063/1.1148867
出版商: AIP
数据来源: AIP
摘要:
The measurement of short distances by interferometry using low coherence light is a potential option in the study of interfaces between intraocular elements. In this article we describe an experimental device to measure the distance between such interfacesin vitro. The procedure utilizes a Michelson interferometer to generate the probe beams together with a silicate photodetector system and a lock-in synchronous amplifier. The cornea and crystalline lens thickness of a rabbit eyeball are evaluated, as well as the human cornea in preservative solution. These tests indicate the method to be efficient, easy-to-handle, of high resolution and to present a signal-to-noise ratio of up to 12. We consider the possibility of its use in obtaining information about the degeneration of intraocular structures during the preservation process. ©1998 American Institute of Physics.
点击下载:
PDF
(805KB)
返 回