首页   按字顺浏览 期刊浏览 卷期浏览 Beam scanning for heavy charged particle radiotherapy
Beam scanning for heavy charged particle radiotherapy

 

作者: Michael Goitein,   George T. Y. Chen,  

 

期刊: Medical Physics  (WILEY Available online 1998)
卷期: Volume 10, issue 6  

页码: 831-840

 

ISSN:0094-2405

 

年代: 1998

 

DOI:10.1118/1.595419

 

出版商: American Association of Physicists in Medicine

 

数据来源: WILEY

 

摘要:

An important aspect of heavy charged particle radiotherapy is its ability to localize dose to the target volume. Current techniques generally employ beam delivery schemes which use range modulated beams in which the Bragg peak is spread out over a range of depths. The range modulation is constant over the entire beam cross section. This paper examines the potential gain from the use of variable range modulation over the beam cross section which can be achieved through beam scanning. An analytic technique has been developed to calculate a gain factor, related to integral dose, for a number of different charged particle beams. The dependence of the gain factor on tumor depth, diameter, and size has been examined. These analytic results have been compared with computer calculations of treatment plans simulating beam scanning and fixed modulation delivery techniques for actual clinical cases. The rationale for beam scanning, and its potential impact on tumor control is discussed.

 

点击下载:  PDF (968KB)



返 回