Continuum x-ray source as a calibration system for charge coupled devices
作者:
K. Hashimotodani,
T. Toneri,
S. Kitamoto,
H. Tsunemi,
K. Hayashida,
E. Miyata,
K. Katayama,
T. Kohomura,
R. Asakura,
K. Koyama,
K. Yamamoto,
K. Miyaguchi,
H. Suzuki,
期刊:
Review of Scientific Instruments
(AIP Available online 1998)
卷期:
Volume 69,
issue 2
页码: 392-395
ISSN:0034-6748
年代: 1998
DOI:10.1063/1.1148671
出版商: AIP
数据来源: AIP
摘要:
We present x-ray spectra obtained by a charge coupled device (CCD) from an electron impacting x-ray source using several kinds of electron targets: Au, Al–bronze, Al, and pure Al. The x rays from the source are dispersed by a grating and the dispersed x rays are focused on the CCD. Owing to the fine spatial and moderate energy resolutions of the CCD, fine spectra with 2 eV energy resolution up to 2.2 keV are obtained. X rays from the pure Al target (99.999&percent; purity) provide good continuum x rays except for AlKx rays. This continuum source is useful as a calibration source for low energy x-ray detectors. ©1998 American Institute of Physics.