首页   按字顺浏览 期刊浏览 卷期浏览 Design and implementation of a heatable, all‐glass doser for sensitive organomet...
Design and implementation of a heatable, all‐glass doser for sensitive organometallic compounds in an ultrahigh vacuum environment

 

作者: L. A. Jones,   F.‐X. Wei,   D. F. Thomas,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1995)
卷期: Volume 66, issue 10  

页码: 4981-4984

 

ISSN:0034-6748

 

年代: 1995

 

DOI:10.1063/1.1146119

 

出版商: AIP

 

数据来源: AIP

 

摘要:

We report the design of an ultrahigh vacuum compatible vapor dosing system for organometallic compounds. A unique feature of this doser is the elimination of all metallic surfaces in order to reduce the reactivity of the organometallic molecules towards the doser walls. The vapors only come in contact with glass, Teflon, and Kalrez and the doser is heatable to temperatures of 150 °C. A novel modification to a commercial glass stopcock is described. These modified valves are used in the doser to seal the ultrahigh vacuum system from the organometallic sample compartment and have been shown to be able to allow the vacuum system to maintain ultrahigh vacuum conditions even when the sample compartment is vented to atmosphere. Spring‐loaded Teflon seals are used to seal the valve shaft against the glass wall thereby maintaining ultrahigh vacuum conditions and a chemically inert surface for the organometallic compounds. A further modification to these commercial valves is described allowing it to function as the sample compartment for the organometallic material. This compartment features a heatable reservoir, vacuum seals to ultrahigh vacuum levels, and a rapid exchange design to permit the interchange of dosing materials. ©1995 American Institute of Physics.

 

点击下载:  PDF (571KB)



返 回