An Electrochemical Determination of Uric Acid in Sera by a Flow-Through Equipment with and Without Uricase
作者:
M. Jänchen,
G. Grünig,
K. Bertermann,
期刊:
Analytical Letters
(Taylor Available online 1985)
卷期:
Volume 18,
issue 15
页码: 1799-1820
ISSN:0003-2719
年代: 1985
DOI:10.1080/00032718508062909
出版商: Taylor & Francis Group
关键词: Uric acid;enzyme electrode;amperometric electrode;flow stream device
数据来源: Taylor
摘要:
A comparison of an enzymeless direct electrochemical oxidation procedure at a platinum electrode for the determination of uric acid, and an enzyme sensor with immobilized urate: oxygen oxidoreductase (uricase), was performed in flow stream systems. The uricase enzyme electrode is based on the H2O2oxidation current. Both amperometric methods were related to the wall-known photometric uricase-catalase-procedure (UCM) as a reference method. The measured values of both methods are of the first derivatives of current change (dI/dt) due to the electrochemical or electrochemical enzymatic reaction, respectively. The analytical quality of the measurements is characterized by: precision s% within run < 2% day to day < 5% accuracy acceptable (control materials) correlation to reference method r >0.93 analysis rate 80 samples/hr
点击下载:
PDF (589KB)
返 回