首页   按字顺浏览 期刊浏览 卷期浏览 The Interpretation Error in Molecular Beam Experiments
The Interpretation Error in Molecular Beam Experiments

 

作者: Andrew Klavins,   Lee H. Sentman,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1970)
卷期: Volume 41, issue 11  

页码: 1560-1567

 

ISSN:0034-6748

 

年代: 1970

 

DOI:10.1063/1.1684341

 

出版商: AIP

 

数据来源: AIP

 

摘要:

This study is concerned with the interpretation of the ionization gauge detector signal in a molecular beam experiment as a particular functional of the reflected molecule distribution function, called the nominal value of the detector signal. The detector configurations examined are those used to measure number density and time‐of‐flight velocity distributions. Expressions for the actual detector signals are established on the basis of the apparatus geometry; the nominal expressions are derived as approximations to the actual signal. Differences between these actual and nominal functionals are examined quantitatively using the Nocilla reflection model. It is found that in the density case the significant apparatus parameter is the solid angle subtended at the detector by the illuminated target area. If this solid angle is large, the error introduced by assuming the detector signal to be represented by the nominal expression becomes significant when the reflection is highly anisotropic. In the velocity distribution measurement, the significant factor is the shutter function. For slow shutter speeds, differences between the actual and nominal expressions for the detector signal increase rapidly as the error in the time the shutter is open increases.

 

点击下载:  PDF (519KB)