Accuracy of methemoglobin measurementsComparison of six different commercial devices and one manual method

 

作者: Jorg,   Dotsch Suha,   Demirakca Dorle,   Hahn Norbert,   Katz Peter Gonne,   Kuhl Wolfgang,  

 

期刊: Critical Care Medicine  (OVID Available online 1999)
卷期: Volume 27, issue 6  

页码: 1191-1194

 

ISSN:0090-3493

 

年代: 1999

 

出版商: OVID

 

数据来源: OVID

 

摘要:

ObjectiveDuring nitric oxide inhalation, methemoglobinemia needs to be monitored. We compared six commercially available instruments and one manual method for methemoglobin measurements. In addition, we studied whether and to what degree methylene blue interferes with methemoglobin measurements.DesignIn vitro methodologic study.SettingResearch laboratory in a university hospital.PatientsFive healthy volunteers from whom red blood cells were obtained.InterventionsMethemoglobinemia was generated in a red blood cell suspension by nitric oxide; methemoglobin was measured with six commercial instruments and one manual photometric method to calculate variation coefficients and to determine the differences between the devices. Methemoglobin was measured with and without the addition of methylene blue with two instruments. Measurements were performed immediately after the addition of methylene blue.Measurements and Main ResultsAll six commercially available instruments had variation coefficients of <0.1 at methemoglobin concentrations of 5%, whereas the manual photometric method did not reach a variation coefficient of <0.1 at 8% of methemoglobin. Apart from two devices that measured slightly but significantly higher methemoglobin levels, all instruments measured similar values of methemoglobin when the same samples were determined simultaneously. Higher concentrations of methylene blue (10, 40, 100 [micro sign]M) reduced substantially the apparent concentrations of methemoglobin. Interference by methylene blue was most pronounced at low methemoglobin levels.ConclusionsWith some limitations, all commercial instruments that were tested performed adequately for the monitoring of methemoglobinemia. Methylene blue interferes with the methemoglobin measurements in a dose-dependent manner. (Crit Care Med 1999; 27:1191-1194)

 



返 回