Flow Injection Determination Of L-Ascorbic Acid in Natural Juice with Biamperometric Detection
作者:
Mihaela CHEREGI,
AndreiF. DANET,
期刊:
Analytical Letters
(Taylor Available online 1997)
卷期:
Volume 30,
issue 14
页码: 2625-2640
ISSN:0003-2719
年代: 1997
DOI:10.1080/00032719708001770
出版商: Taylor & Francis Group
关键词: Flow-Injection System;Biamperometric Detection;L-Ascorbic Acid (Vitamin C) Determination;Ascorbate oxidase;Orange Natural Juice
数据来源: Taylor
摘要:
A flow-injection system with biamperometric detection for L-ascorbic acid (vitamin C) determination was designed. The method is based on L-ascorbic acid oxidation to dehydroascorbic acid, in acidic medium, using iodine-iodide solution as oxidizing reagent. The iodine amount consumed in the redox reaction was biaperometrically detected and it was proportional to the amount of L-ascorbic acid from the sample. The calibration graph was linear over the range of concentration 5×10−5-5×10−4M of vitamin C. The RSD for a 2.5×10−4M standard solution of vitamin C was 1.08% (n=10) and the throughput was 60 samples h−1. The selectivity of the proposed FIA method was verified using an ascorbate oxidase solution as carrier stream. It was found that L-ascorbic acid could be determined in natural products with a good selectivity and rapidity.