首页   按字顺浏览 期刊浏览 卷期浏览 High energy x‐ray imaging diagnostic on Nova
High energy x‐ray imaging diagnostic on Nova

 

作者: B. A. Remington,   B. A. Hammel,   O. L. Landen,   R. A. Pasha,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1992)
卷期: Volume 63, issue 10  

页码: 5083-5085

 

ISSN:0034-6748

 

年代: 1992

 

DOI:10.1063/1.1143499

 

出版商: AIP

 

数据来源: AIP

 

摘要:

We have developed a diagnostic suitable for high photon energy (∼7 keV) time resolved one‐dimensional x‐ray imaging on Nova. This diagnostic uses a 12 &mgr;m wide, 750 &mgr;m long imaging slit mounted on a 20× magnification snout which is coupled to an x‐ray streak camera of variable streak speed. The diagnostic fits into retractable manipulators on the ten‐beam chamber at Nova.Insitubacklit grid experiments using a large‐area backlighter give a spatial resolution full width at half‐maximum of about 15 &mgr;m with high brightness.  

 