首页   按字顺浏览 期刊浏览 卷期浏览 Count rate performance of spectrometers for fusion neutrons
Count rate performance of spectrometers for fusion neutrons

 

作者: J. Ka¨llne,   G. Gorini,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1993)
卷期: Volume 64, issue 10  

页码: 2765-2770

 

ISSN:0034-6748

 

年代: 1993

 

DOI:10.1063/1.1144415

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The count rate capability of high‐resolution neutron spectrometers for fusion plasmas is discussed. In particular, the question of the intrinsic rate limitations in neutron time‐of‐flight (TOF) spectrometers is assessed quantitatively for two design approaches where the proton recoils fromn+Hscatterings are detected with in‐beam and out‐of‐beam detectors. It is found that the count rate capability limit lies in the range 50–400 kHz for TOF depending on approach. Comparison is made with other types of neutron spectrometers for fusion experiments. It is shown, for the first time, that neutron spectrometers fall into distinct categories depending on the underlying detection methods and that this information can be used for assessing capability limits of neutron spectrometry as a diagnostic for fusion plasmas.

 

点击下载:  PDF (931KB)