11. |
Displacement‐Type He II Heat Switch |
|
Review of Scientific Instruments,
Volume 31,
Issue 9,
1960,
Page 966-969
Ferdinand J. Shore,
Preview
|
PDF (285KB)
|
|
摘要:
The operation of a heat switch is described in which the conducting substance is liquid He II and in which the nonconducting state is achieved by displacing the He II with a poor conductor, e.g., Teflon. The switch is well suited for use in the first stage of a magnetic refrigerator operating with a bath temperature near 1°K. At this temperature it performs better than a lead superconducting switch.
ISSN:0034-6748
DOI:10.1063/1.1717119
出版商:AIP
年代:1960
数据来源: AIP
|
12. |
Use of Graphite as Low Temperature Support and Shunt for Heat Switch |
|
Review of Scientific Instruments,
Volume 31,
Issue 9,
1960,
Page 970-973
F. J. Shore,
V. L. Sailor,
H. Marshak,
C. A. Reynolds,
Preview
|
PDF (307KB)
|
|
摘要:
Artificial pitch‐bonded graphite is known to exhibit an extremely large variation of heat conductivity with temperature. It is both rigid and strong, and below 1°K its heat conductivity is exceptionally small. For these reasons we have found it to be an ideal support material for use within the adiabatic enclosure of a cryostat operated below 1°K. Furthermore, above liquid nitrogen temperature graphite is an excellent heat conductor, and when used in parallel with a lead heat switch, it proves unnecessary to admit exchange gas during the precool period. The graphite exhibits a strong pumping action for helium gas at low temperatures.
ISSN:0034-6748
DOI:10.1063/1.1717120
出版商:AIP
年代:1960
数据来源: AIP
|
13. |
Effect of Crystal Thickness and Geometry on the Alpha‐Particle Resolution of CsI (Tl) |
|
Review of Scientific Instruments,
Volume 31,
Issue 9,
1960,
Page 974-977
Prudencio Martinez,
Frank E. Senftle,
Preview
|
PDF (302KB)
|
|
摘要:
The resolution of CsI(Tl) for Po210alpha particles has been measured as a function of crystal thickness. The best resolution of a ½‐in. diam cylindrical crystal was obtained for a thickness of 0.38 mm, and the effect of thickness on the resolution is discussed. Based on the proposed model, a conical crystal was designed, which yielded a line width of 1.8% for Po210alpha particles with a selected photomultiplier tube.
ISSN:0034-6748
DOI:10.1063/1.1717121
出版商:AIP
年代:1960
数据来源: AIP
|
14. |
Simple Transistor Marginal Oscillator for Magnetic Resonance |
|
Review of Scientific Instruments,
Volume 31,
Issue 9,
1960,
Page 977-978
Bailey Donnally,
T. M. Sanders,
Preview
|
PDF (135KB)
|
|
摘要:
A transistor marginal oscillator circuit for nuclear magnetic resonance measurements is described. The circuit is self‐contained, simple, nonmicrophonic, and noncritical and has been used in a variety of measurements in both research and instructional laboratories. The sensitivity of the device is approximately one‐half that of a Pound‐Knight‐Watkins system.
ISSN:0034-6748
DOI:10.1063/1.1717122
出版商:AIP
年代:1960
数据来源: AIP
|
15. |
Ultralinear Cathode Follower |
|
Review of Scientific Instruments,
Volume 31,
Issue 9,
1960,
Page 979-982
Philip L. Read,
Preview
|
PDF (268KB)
|
|
摘要:
The design of a modified cathode follower circuit which possesses an extremely linear response and a voltage gain of essentially unity is presented. Analysis of the circuit shows that, in principle, the harmonic distortion may be made arbitrarily small, and at the same time the voltage gain may be made arbitrarily close to unity. The modification is applied to a White cathode follower where the increase in linearity and gain is accompanied by a decrease in the output impedance. A practical modified White follower having a measured intermodulation distortion of 2×10−5%, a calculated voltage gain of (1−10−6), and a calculated output impedance of 2×10−3ohms is given.
ISSN:0034-6748
DOI:10.1063/1.1717123
出版商:AIP
年代:1960
数据来源: AIP
|
16. |
Six‐Meter Radius Bent‐Crystal Spectrograph for Nuclear Gamma Rays |
|
Review of Scientific Instruments,
Volume 31,
Issue 9,
1960,
Page 983-987
A. H. Kazi,
N. C. Rasmussen,
Hans Mark,
Preview
|
PDF (379KB)
|
|
摘要:
A six‐meter radius bent‐crystal spectrograph for the precision measurement of nuclear gamma‐ray wavelengths has been designed, constructed, and placed into operation at the Massachusetts Institute of Technology. This instrument is particularly well suited for directly measuring neutron capture gamma rays produced by samples placed in the through port of the MIT research reactor. The spectrograph is designed to measure wavelengths of gamma rays in the energy region from 0.1 to 4.0 Mev with an estimated precision that varies from about 0.01% at 0.1 Mev to 0.3% at 4 Mev. Photographic methods are used to record the gamma‐ray lines on the focal circle. A bent crystal is used to diffract and focus the gamma rays. Line width and efficiency data have been obtained for the (310), (003), and (006) planes of quartz. The efficiency of the instrument is quite low. In order to record a line at 2.0 Mev an exposure time of about 6000 curie hours is required.
ISSN:0034-6748
DOI:10.1063/1.1717124
出版商:AIP
年代:1960
数据来源: AIP
|
17. |
Intracardiac Catheter Tip Piezoresistive Pressure Gauge |
|
Review of Scientific Instruments,
Volume 31,
Issue 9,
1960,
Page 987-991
Max Traite,
Walter Welkowitz,
Rudolph Downs,
Preview
|
PDF (366KB)
|
|
摘要:
A pressure gauge is described that can be encased in a small‐bore plastic cathether. The small size permits insertion directly into the heart. The unit uses a two‐section piezoresistive element, operated as a cantilever. The sensitivity of the gauge is 0.16 mv/v/psi, and the frequency response is flat from 0 to 400 cps.
ISSN:0034-6748
DOI:10.1063/1.1717125
出版商:AIP
年代:1960
数据来源: AIP
|
18. |
Transparent Indium Contacts to CdS |
|
Review of Scientific Instruments,
Volume 31,
Issue 9,
1960,
Page 992-994
Y. T. Sihvonen,
D. R. Boyd,
Preview
|
PDF (239KB)
|
|
摘要:
Diffusion and sputtering techniques are described for applying low resistance, highly transparent, antireflecting, electrical contacts to CdS. The diffusion process entails heating CdS for several minutes in the presence of indium vapor. The resulting crystal surface resistance has been adjusted to range as low as 12 ohms/square without any observable crystal discoloration. The sputtering process employs an 82–18% indium‐tin alloy that is placed onto the crystal surface using oxygen as the sputtering atmosphere. The resulting film is Sn‐doped In2O3, measuring 100–150 ohms/square, and showing 80–100% transmittance from 4600 to 20 000 A.
ISSN:0034-6748
DOI:10.1063/1.1717126
出版商:AIP
年代:1960
数据来源: AIP
|
19. |
Fast Magnetic Field Sweep |
|
Review of Scientific Instruments,
Volume 31,
Issue 9,
1960,
Page 995-997
G. W. Barton,
L. F. Tolman,
R. E. Roulette,
Preview
|
|
ISSN:0034-6748
DOI:10.1063/1.1717128
出版商:AIP
年代:1960
数据来源: AIP
|
20. |
Transistorized Lock‐In for Klystron Automatic Frequency Control |
|
Review of Scientific Instruments,
Volume 31,
Issue 9,
1960,
Page 997-998
A. J. George,
D. T. Teaney,
Preview
|
|
ISSN:0034-6748
DOI:10.1063/1.1717130
出版商:AIP
年代:1960
数据来源: AIP
|
|