21. |
High‐performance, 2.5‐K cryostat incorporating a 100–120‐GHz dual polarization receiver |
|
Review of Scientific Instruments,
Volume 56,
Issue 3,
1985,
Page 449-458
John W. Archer,
Preview
|
PDF (792KB)
|
|
摘要:
A hybrid cryostat is described which was designed to cool two SIS mixers in a 100–120‐GHz dual polarization receiver. A mechanical refrigerator is used to cool radiation shields surrounding a 4.3‐l capacity metal tank containing liquid helium (He4). The temperature of the liquid can be lowered to 2.5 K by reducing the helium vapor pressure in the tank with the aid of a mechanical vacuum pump. A liquid hold time, at 2.5 K, of longer than five days has been achieved. With the mixers cooled to 2.5 K andL‐band GaAs FET IF amplifiers cooled to 20 K, single sideband receiver temperatures of less than 120 K have been measured, with >15‐dB image rejection and a useful instantaneous bandwidth of 500 MHz.
ISSN:0034-6748
DOI:10.1063/1.1138321
出版商:AIP
年代:1985
数据来源: AIP
|
22. |
High‐resolution pressure–volume measurements by a strain gauge technique |
|
Review of Scientific Instruments,
Volume 56,
Issue 3,
1985,
Page 459-461
B. Batlogg,
H. Boppart,
Preview
|
PDF (207KB)
|
|
摘要:
An experimental technique is described which employs resistive strain gauges to measure the pressure–volume relationship of solids up to∼50 kbar. The strain resolution is better than 10 ppm with a dynamical range up to 20 000. Two examples are presented to illustrate the sensitivity and the reliability of this technique.
ISSN:0034-6748
DOI:10.1063/1.1138322
出版商:AIP
年代:1985
数据来源: AIP
|
23. |
Automation of an alternating field microelectrophoresis apparatus |
|
Review of Scientific Instruments,
Volume 56,
Issue 3,
1985,
Page 462-465
B. H. Hakim,
E. Delatour,
M. Hanss,
Preview
|
PDF (306KB)
|
|
摘要:
The automation of a microelectrophoresis apparatus is described. It relies on digital techniques for the electric field generation, the conductance measurement, the cuvette positioning, and the temperature rise estimation deduced from the conductance. A current source is used with programmable shape, frequency, duty cycle, and duration. All the measurement steps are successively controlled by the microprocessor. As described herein, the particles motion is recorded with a camera. The apparatus is built from independent (software and hardware) modules and other detecting devices could easily be adapted.
ISSN:0034-6748
DOI:10.1063/1.1138323
出版商:AIP
年代:1985
数据来源: AIP
|
24. |
Low‐frequency digital lock‐in amplifier |
|
Review of Scientific Instruments,
Volume 56,
Issue 3,
1985,
Page 466-470
P. A. Probst,
B. Collet,
Preview
|
PDF (362KB)
|
|
摘要:
A digital lock‐in amplifier which simultaneously measures in phase and in quadrature components at a reference frequency and at twice this frequency is described. It is drift free, insensitive to harmonics, has a very high gain stability (25 ppm/K), and operates up to 1250 Hz with no limitations on the low‐frequency side.
ISSN:0034-6748
DOI:10.1063/1.1138324
出版商:AIP
年代:1985
数据来源: AIP
|
25. |
Solid neon sources for plasmas and x‐ray lasers |
|
Review of Scientific Instruments,
Volume 56,
Issue 3,
1985,
Page 471-472
R. H. Dixon,
J. L. Ford,
T. N. Lee,
R. C. Elton,
Preview
|
PDF (151KB)
|
|
摘要:
A thin layer of frozen neon is condensed onto a cryogenically cooled substrate for use in generating a neon plasma for x‐ray laser research. The freezing process and quasiequilibrium state of the layer are discussed. Spectral purity of a neon plasma formed by laser irradiation has been demonstrated. Control of the layer thinness is possible.
ISSN:0034-6748
DOI:10.1063/1.1138526
出版商:AIP
年代:1985
数据来源: AIP
|
26. |
Irradiation cell for electron pulse radiolysis in liquid anhydrous hydrogen fluoride and other corrosive fluoride solvents |
|
Review of Scientific Instruments,
Volume 56,
Issue 3,
1985,
Page 473-474
D. Martin‐Rovet,
B. Hickel,
R. Guitton,
J. Potier,
Preview
|
PDF (184KB)
|
|
摘要:
An irradiation cell for electron pulse radiolysis experiments in reactive liquid fluoride media, such as anhydrous hydrogen fluoride, is described. The construction material is Kel‐F, the irradiation window being a part of the body of the cell. The optical windows for the analytical light are made from sapphire. The cell has been tested with reactive solutions which have been added and removed on a vacuum line, and has been shown to be suitable for electron beam radiolysis studies.
ISSN:0034-6748
DOI:10.1063/1.1138325
出版商:AIP
年代:1985
数据来源: AIP
|
27. |
Simple circuit to passively synchronize the predischarge and main discharge in a UV preionized TEA CO2laser |
|
Review of Scientific Instruments,
Volume 56,
Issue 3,
1985,
Page 474-476
N. S. Shikarkhane,
U. Nundy,
U. K. Chatterjee,
Preview
|
PDF (256KB)
|
|
摘要:
A simple, inexpensive circuit for synchronizing the preionizer and the main discharge in an UV preionized TEA CO2laser is described. The circuit uses only two coupling capacitors of value 1.35 nF, between the two discharge circuits to achieve the synchronization. For a delay of 1 &mgr;s, the jitter is ±50 ns.
ISSN:0034-6748
DOI:10.1063/1.1138270
出版商:AIP
年代:1985
数据来源: AIP
|
28. |
Using the electronic top‐loading analytical balance for surface tension measurements |
|
Review of Scientific Instruments,
Volume 56,
Issue 3,
1985,
Page 476-477
Jack Richlin,
Preview
|
PDF (164KB)
|
|
摘要:
An electronic top‐loading analytical balance has been adapted to measure surface tension at liquid–gas interfaces. Because the zero position of the balance is automatically returned, the instrument is capable of good accuracy.
ISSN:0034-6748
DOI:10.1063/1.1138271
出版商:AIP
年代:1985
数据来源: AIP
|
29. |
High‐accuracy optical attenuator continuously variable in the range of 1:20 |
|
Review of Scientific Instruments,
Volume 56,
Issue 3,
1985,
Page 478-479
Mikio Yamashita,
Yasushi Kawada,
Satoshi Takeuchi,
Preview
|
PDF (167KB)
|
|
摘要:
A continuously variable optical attenuator is described that uses optical fiber light guides and high‐accuracy dial micrometers mounted on mechanical stages. The attenuation factors thereby obtained have proved reproducible to ∼0.1% over a dynamic range of 1:20. It should, therefore, be particularly useful in testing the linearity of the response of various photodetectors.
ISSN:0034-6748
DOI:10.1063/1.1138272
出版商:AIP
年代:1985
数据来源: AIP
|
30. |
Pulsed nuclear‐magnetic‐resonance thermometer with a microprocessor for ultralow temperatures |
|
Review of Scientific Instruments,
Volume 56,
Issue 3,
1985,
Page 480-481
Anju Sawada,
Hideo Yano,
Yoshika Masuda,
Preview
|
PDF (182KB)
|
|
摘要:
A pulsed nuclear‐magnetic‐resonance (NMR) thermometer with platinum as a probe, which can operate at the NMR frequencies of 62.5, 125, 250, and 500 kHz, has been constructed. All of the various pulsed NMR sequences can be executed by commands from a microprocessor. The present system has excellent performance characteristics such as long‐term stability, high resolution (16 bits), full automatic operation, and simplicity in electronic circuitry. It has been used at temperatures as low as 0.45 mK.
ISSN:0034-6748
DOI:10.1063/1.1138273
出版商:AIP
年代:1985
数据来源: AIP
|
|