Modified airlock door for the introduction of frozen specimens into the JEM 100B electron microscope
作者:
Kenneth A. Taylor,
Robert M. Glaeser,
期刊:
Review of Scientific Instruments
(AIP Available online 1975)
卷期:
Volume 46,
issue 8
页码: 985-986
ISSN:0034-6748
年代: 1975
DOI:10.1063/1.1134383
出版商: AIP
数据来源: AIP
摘要:
An airlock door modification for the JEM 100B electron microscope is described. This modification permits prefrozen specimens to be introduced into the microscope without frost forming on the specimen or the specimen warming excessively. The imaging of hydrated biological material using frozen specimens is demonstrated with crystalline catalase as a test specimen.