Hemodynamic Comparisons of Polyurethane Trileaflet and Bioprosthetic Heart Valves
作者:
Krishnan Chandran,
Richard Schoephoerster,
David Wurzel,
Gary Hansen,
Long Yu,
George Pantalos,
Willem Kolff,
期刊:
ASAIO Journal
(OVID Available online 1989)
卷期:
Volume 35,
issue 2
页码: 132-138
ISSN:1058-2916
年代: 1989
出版商: OVID
数据来源: OVID
摘要:
&NA;Hemodynamic comparison of two polyurethane prosthetic heart valves with a bioprosthetic valve is presented. The valves were incorporated in a pulse duplicator simulating physiologic pulsatile flow, and comparisons between the valves were made on the transvalvular pressure drop, percent regurgitation, valve orifice area, rate of opening and closing as well as the performance index. The results showed that the functional characteristics of the polyurethane valves compared favorably with that of the bioprosthetic valve. The polyurethane valve can be a viable and inexpensive alternative, especially for short‐term use in a total artificial heart as a bridge to transplant.ASAIO Transactions1989; 35: 132–138.