首页   按字顺浏览 期刊浏览 卷期浏览 Pitfalls of Swan‐Ganz catheterization
Pitfalls of Swan‐Ganz catheterization

 

作者: BAEKHYO SHIN,   ROBERT AYELLA,   T. McASLAN,  

 

期刊: Critical Care Medicine  (OVID Available online 1977)
卷期: Volume 5, issue 3  

页码: 125-127

 

ISSN:0090-3493

 

年代: 1977

 

出版商: OVID

 

数据来源: OVID

 

摘要:

In 60 patients in whom Swan-Ganz catheters apparently had been positioned correctly, the balloon was visualized by inflation with radiopaque contrast medium. Sixteen were located peripherally; in 15 of these 16, the balloon inflated eccentrically and in each of these instances, an accurate wedge pressure could not be obtained. One patient in this group had an episode of hemoptysis immediately prior to detection of the peripheral location and eccentric inflation of the balloon.The correct placement and safe use of the Swan-Ganz catheter demand that the catheter tip be located in a large pulmonary artery and that redundant loops of catheter be avoided to prevent subsequent peripheral migration. Identification of peripheral placement and eccentric inflation should be suspected if a pulmonary wedge pressure is obtained with a significantly smaller volume of air than the balloon capacity. The use of a continuous flush system will provide an additional alert by a steady rise in the pseudowedge pressure on attempted balloon inflation.

 

点击下载:  PDF (224KB)