首页   按字顺浏览 期刊浏览 卷期浏览 Sterilization with gaseous ethylene oxide: A review of chemical and physical factors
Sterilization with gaseous ethylene oxide: A review of chemical and physical factors

 

作者: Robert R. Ernst,   John E. Doyle,  

 

期刊: Biotechnology and Bioengineering  (WILEY Available online 1968)
卷期: Volume 10, issue 1  

页码: 1-31

 

ISSN:0006-3592

 

年代: 1968

 

DOI:10.1002/bit.260100102

 

出版商: Wiley Subscription Services, Inc., A Wiley Company

 

数据来源: WILEY

 

摘要:

AbstractAlthough the basic parameters for ethylene oxide sterilization are established, it is sometimes difficult to attain in practice where the principal limiting factor is moisture availability. There are situations which can limit or enhance the dynamics of sterilization. Such factors, if overlooked, could upset experiments and lead to erroneous conclusions, or defeat the sterilization process entirety. Such are, namely: stratification effects, diffusion barriers, moisture‐reducing effects, polymerization, and temperature distribution gradient

 

点击下载:  PDF (1644KB)



返 回