首页   按字顺浏览 期刊浏览 卷期浏览 Precision isoperibol calorimeter with automatic data acquisition and processing
Precision isoperibol calorimeter with automatic data acquisition and processing

 

作者: C. Shin,   C. M. Criss,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1975)
卷期: Volume 46, issue 8  

页码: 1043-1046

 

ISSN:0034-6748

 

年代: 1975

 

DOI:10.1063/1.1134390

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A precision isoperibol calorimeter with automatic data acquisition and processing, capable of operating over a wide temperature range, has been constructed and tested. The calorimeter was designed to have a very low thermal modulus and fast equilibration time. A small vapor space above the liquid and a vapor trap make the calorimeter useful for measurements with nonaqueous solvents. Time–temperature data are automatically recorded in digital form on magnetic tape and processed on a Univac 1106 computer. A series of electrical calibrations causing a temperature rise of 0.13 K were reproducible to about 0.01%. Several determinations of the enthalpy of reaction of a standard sample of Tris (hydroxymethyl) aminomethane with 0.1NHCl gave an average value of 29.764±0.013 kJ/mol, which agrees satisfactorily with published results.

 

点击下载:  PDF (315KB)



返 回