Low‐temperature thermometry in high magnetic fields. VI. Industrial‐grade Pt resistors above 66 K; Rh–Fe and Au–Mn resistors above 40 K
作者:
B. L. Brandt,
L. G. Rubin,
H. H. Sample,
期刊:
Review of Scientific Instruments
(AIP Available online 1988)
卷期:
Volume 59,
issue 4
页码: 642-645
ISSN:0034-6748
年代: 1988
DOI:10.1063/1.1139849
出版商: AIP
数据来源: AIP
摘要:
The characteristics of twelve industrial‐grade and one precision platinum resistance thermometers for temperatures from 66 to 300 K and in magnetic fields up to 19 T are described. The results from the 13 models from nine suppliers are combined to produce mean values that may be used to correct any of those thermometers for effects of a magnetic field parallel to its long axis. The correctability error at 19 T is less than ±0.3% between 66 and 110 K and less than ±0.1% above 110 K. Magnetoresistance data for Rh–Fe and Au–Mn resistance thermometers are also presented.