Effektivität unterschiedlicher Tonometer-Desinfektionsverfahren
作者:
A. Wizemann,
期刊:
Ophthalmologica
(Karger Available online 1980)
卷期:
Volume 181,
issue 1
页码: 13-22
ISSN:0030-3755
年代: 1980
DOI:10.1159/000309020
出版商: S. Karger AG
关键词: Effectivity of desinfection;Desinfection of tonometer
数据来源: Karger
摘要:
In 13,440 bacteriological investigations the effectivity of different methods for desinfection of Goldmann tonometer prisms (merfen, pantasept, sekusept steril, UV sterilizer Sklar) is tested. Sekusept steril and UV sterilizer show the quickest and most sufficient results. They cause no damage to corneal epithelium or to plastic objects. We can recommend them for desinfection of applanation tonometer prims, Goldmann’s gonioscopy and fundus contact lenses as routine methods.
点击下载:
PDF
(2866KB)
返 回