首页   按字顺浏览 期刊浏览 卷期浏览 Predicting total body water and extracellular fluid volumes from bioelectrical measurem...
Predicting total body water and extracellular fluid volumes from bioelectrical measurements of the human body.

 

作者: JohnsonH L,   VirkS P,   MayclinP,   BarbieriT,  

 

期刊: Journal of the American College of Nutrition  (Taylor Available online 1992)
卷期: Volume 11, issue 5  

页码: 539-547

 

ISSN:0731-5724

 

年代: 1992

 

DOI:10.1080/07315724.1992.10718259

 

出版商: Routledge

 

数据来源: Taylor

 

摘要:

Two biological impedance analyzers, a 50 kHz (RJL) and 20-100 kHz (BMA) instrument, and a total body electrical conductivity (TOBEC) instrument were used to estimate total body water (TBW), extracellular (ECF) and intracellular (ICF) fluid volumes by repeated measurements of 16 normal men (19-38 years old) to assess which, if any, would provide the best estimates. At 3-week intervals, TBW was determined by deuterium dilution, ECF by bromide dilution, ICF by difference (TBW-ECF) and lean body mass by density. Prediction equations were obtained by regression; predicted values for the body fluid volumes were calculated and the results were statistically evaluated. Both the TOBEC and the BMA provided rapid and reliable estimates for body fluid volumes with standard errors of the estimates of about 0.5-1.1 L for ECF, 1.0-1.8 L for TBW, and 1.0-1.3 L for ICF. Part of the error was attributable to standard tracer-dilution methods.

 

点击下载:  PDF (744KB)



返 回