首页   按字顺浏览 期刊浏览 卷期浏览 Monitoring Seizure Duration During Electroconvulsive Therapy
Monitoring Seizure Duration During Electroconvulsive Therapy

 

作者: Lawrence Couture,   Linda Lucas,   Steven Lippmann,   Taher Shaltout,   Markku Paloheimo,   Harvey Edmonds,  

 

期刊: Convulsive Therapy  (OVID Available online 1988)
卷期: Volume 4, issue 3  

页码: 206-214

 

ISSN:0749-8055

 

年代: 1988

 

出版商: OVID

 

关键词: Electroconvulsive therapy;Electroencephalogram;Electromyogram;Seizure duration;Quantitative electroencephalogram

 

数据来源: OVID

 

摘要:

Three alternative monitoring methods for assessing the duration of seizures during electroconvulsive therapy were studied. The mean integrated amplitude of the electroencephalogram, facial muscle electromyogram, and “cuff method” were compared with the reference single-channel unprocessed electroencephalogram in 78 sessions with 17 patients. The measures of seizure duration differed significantly (p< 0.001). The mean integrated electroencephalographic amplitude differences were small and an artifact of the sampling procedure. Larger, but clinically unimportant, discrepancies were obtained with the facial electromyogram amplitude. In contrast, there were marked differences between the electroencephalogram and the “cuff method,” which suggest that the latter technique may be of limited usefulness.

 