首页   按字顺浏览 期刊浏览 卷期浏览 Relatively Simple Modification of an AEI MS-902 High-Resolution Mass Spectrometer to Pe...
Relatively Simple Modification of an AEI MS-902 High-Resolution Mass Spectrometer to Permit Chemical Ionization Studies

 

作者: W. A. Garland,   R. J. Weinkam,   W. F. Trager,  

 

期刊: Instrumentation Science & Technology  (Taylor Available online 1974)
卷期: Volume 5, issue 4  

页码: 271-281

 

ISSN:1073-9149

 

年代: 1974

 

DOI:10.1080/10739147308543377

 

出版商: Taylor & Francis Group

 

数据来源: Taylor

 

摘要:

Relatively inexpensive and simple modifications to an AEI MS-902 double focusing high-resolution mass spectrometer allow chemical ionization studies to be undertaken without compromising the instrument's high resolving power or sensitivity. A standard AEI source block was made gas tight except for a 0.013 to 0.020 in. diameter ion exit and an 0.013-in. electron entrance hole. A specially designed direct insertion probe was utilized to introduce both the sample and reagent gas. The reagent gas itself was supplied via a system utilizing both the manufacturers cold inlet system and Matheson lecture bottles. An additional pumping system was added in parallel to the existing pumping system to handle the increased gas flow.

 