A linear, liquid‐resistivity meter circuit is described that operates by charging and discharging a capacitor through a cell in which the liquid is included, rendering no dc component across it. The resulting oscillating waveform is shown to have a period proportional to the cell resistance. A period–voltage converter is added in order to indicate the resistance on a moving coil instrument. The system’s dc supply and low consumption permit easy battery operation.