首页   按字顺浏览 期刊浏览 卷期浏览 Hazard to Endotracheal Tubes by CO2Laser Beam
Hazard to Endotracheal Tubes by CO2Laser Beam

 

作者: Naoki Ohashi,   Masatugu Asai,   Shinsuke Ueda,   Jun-ichi Imamura,   Yukio Watanabe,   Kanemasa Mizukoshi,  

 

期刊: ORL  (Karger Available online 1985)
卷期: Volume 47, issue 1  

页码: 22-25

 

ISSN:0301-1569

 

年代: 1985

 

DOI:10.1159/000275740

 

出版商: S. Karger AG

 

关键词: CO2laser;Microlaryngo surgery

 

数据来源: Karger

 

摘要:

Ignition or combustion of the endotracheal tube during CO2 laser microlaryngo surgery may cause fatal lung disorders. We undertook an experiment on ignition and combustion of endotracheal tubes by CO2 laser beam and came to the following conclusions: (a) the silicon endotracheal tube was nonflammable; (b) the spiral endotracheal tube was also nonflammable; (c) a pulse wave was recommended and 3-second intervals were necessary; (d) four sheets of moist gauze were necessary to prevent cuff trouble, and (e) it was advisable to set the O2 concentration below 25 % when using the silicon endotracheal tube.

 