Computer‐Assisted Eye Examination VI. Identification and Correction of Errors in the Refractor III System for Subjective Examination
作者:
ELWIN MARG,
KENNETH ANDERSON,
KHIN CHUNG,
CHACKO NEROTH,
期刊:
American Journal of Optometry and Physiological Optics
(OVID Available online 1978)
卷期:
Volume 55,
issue 4
页码: 249-266
ISSN:0093-7002
年代: 1978
出版商: OVID
关键词: refractive error;subjective examination;computer assistance;measurement errors;prescription evaluation
数据来源: OVID
摘要:
A computer-assisted subjective eye examination was administered to 80 patients. They were also examined in the conventional manual way. The results of the computer-assisted examination were compared with those obtained by conventional subjective methods to identify errors of hardware, software, and optometric flow charting and other problems. An 87.5% satisfactory prescription level was found for the distance prescription. The 12.5% unsatisfactory results are accounted for by several kinds of error. The percentage of unsatisfactory refractive results was greater for near than for distance measures because of special flow-chart and software problems, which were subsequently corrected. With implementation of the changes initiated as a result of this investigation, satisfactory refractive results approaching 95% can be expected.