首页   按字顺浏览 期刊浏览 卷期浏览 Multichannel boxcar‐averaged measurements of plasma parameters made using a digi...
Multichannel boxcar‐averaged measurements of plasma parameters made using a digital storage scope

 

作者: T. E. Sheridan,   M. A. Hayes,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1988)
卷期: Volume 59, issue 7  

页码: 1081-1084

 

ISSN:0034-6748

 

年代: 1988

 

DOI:10.1063/1.1139730

 

出版商: AIP

 

数据来源: AIP

 

摘要:

We present a technique for rapidly acquiring time‐resolved, ensemble‐averaged Langmuir probe characteristics. Fifty probe characteristics are acquired using a digital storage oscilloscope in the time it would take to acquire a one‐probe characteristic using a single‐channel boxcar averager. A single Langmuir probe is used, and the probe bias is swept quite slowly, so that the probe is always in equilibrium with the plasma. A method for the automatic extraction of electron temperature, electron density, and the plasma potential from the acquired probe characteristics is described. This technique for acquisition and analysis is applied to the study of plasma decay and the effects of rf excitation in a pulsed, strongly magnetized plasma.

 