The Bacteriological Quality of Hemodialysis Solution as Related to Several Environmental Factors
作者:
James Lauer,
Andrew Streifel,
Carl Kjellstrand,
Roger DeRoos,
期刊:
Nephron
(Karger Available online 1975)
卷期:
Volume 15,
issue 2
页码: 87-97
ISSN:1660-8151
年代: 1975
DOI:10.1159/000180500
出版商: S. Karger AG
关键词: Hemodialysis;Reverse osmosis;Febrile reactions;Artificial kidney machines;Endotoxins;Bacterial contamination
数据来源: Karger
摘要:
The bacterial concentrations of the municipal water increased by more than 39-fold when subjected to reverse osmosis; then decreased by greater than 200-fold within the reservoir and water supply system of the hemodialysis center. The bacterial concentrations of dialysate solutions in contact with proportioning single-pass artificial kidney machines were as low or lower than the water from the hemodialysis center system ( < 10 CFU/100 ml.). The complete opposite was observed in the recirculating single-pass artificial kidney machines where bacterial concentrations in the dialysate solution reached levels greater than 1.0 × 106 CFU/100 ml