The Bacteriological Quality of Hemodialysis Solution as Related to Several Environmental Factors

 

作者: James Lauer,   Andrew Streifel,   Carl Kjellstrand,   Roger DeRoos,  

 

期刊: Nephron  (Karger Available online 1975)
卷期: Volume 15, issue 2  

页码: 87-97

 

ISSN:1660-8151

 

年代: 1975

 

DOI:10.1159/000180500

 

出版商: S. Karger AG

 

关键词: Hemodialysis;Reverse osmosis;Febrile reactions;Artificial kidney machines;Endotoxins;Bacterial contamination

 

数据来源: Karger

 

摘要:

The bacterial concentrations of the municipal water increased by more than 39-fold when subjected to reverse osmosis; then decreased by greater than 200-fold within the reservoir and water supply system of the hemodialysis center. The bacterial concentrations of dialysate solutions in contact with proportioning single-pass artificial kidney machines were as low or lower than the water from the hemodialysis center system ( < 10 CFU/100 ml.). The complete opposite was observed in the recirculating single-pass artificial kidney machines where bacterial concentrations in the dialysate solution reached levels greater than 1.0 × 106 CFU/100 ml

 