首页   按字顺浏览 期刊浏览 卷期浏览 Evaluation and use of the white blood cell differential provided by the Coulter® STKS i...
Evaluation and use of the white blood cell differential provided by the Coulter® STKS in a children's hospital

 

作者: M. FOURNIER,   C. ADENIS,   H. FONTAINE,   B. CARNAILLE,   J. GOUDEMAND,  

 

期刊: Clinical&Laboratory Haematology  (WILEY Available online 1994)
卷期: Volume 16, issue 1  

页码: 33-42

 

ISSN:0141-9854

 

年代: 1994

 

DOI:10.1111/j.1365-2257.1994.tb00385.x

 

出版商: Blackwell Publishing Ltd

 

关键词: WBC Diff;Flow Cytometry;Coulter® STKS;paediatric use

 

数据来源: WILEY

 

摘要:

Summary.The Coulter® STKS was evaluated in a children's hospital, in order to (a) compare theWBCdifferential given by the instrument to a 400 cell visual differential (reference method); (b) evaluate the sensitivity and specificity of the alarm system, and (c) provide data concerning the use and interpretation of results in children. 653 blood samples were collected. The Coulter® STKS results were studied in 523 patients having no morphological abnormalities in the blood smears, separated into subgroups according to the presence of STKS alarms and according to age. The results were found accurate both in STKS negative and STKS positive patients (i.e., those with alarms:‘Blasts', Imm Gran 2, Variant Lymph, NRBC, review slide). Negative STKS results had the same accuracy in all age groups, except in neonates where slide review must be systematically performed. The instrument exhibited a good sensitivity of the suspect flags studied (91.4%), with a lower specificity (72%) reflecting the number of false positive results found in our group, probably due to the cytological features particular to children. However, it was shown that the numerical results given by the Coulter® STKS in positive patients could be taken into account, provided that a scan of the blood smear was negative for morphological WBC abnormali

 

点击下载:  PDF (545KB)



返 回