Highly Stable High‐Rate Discriminator for General Use in Nuclear Counting Applications
作者:
S. J. Rudnick,
J. J. English,
R. H. Howard,
期刊:
Review of Scientific Instruments
(AIP Available online 1969)
卷期:
Volume 40,
issue 1
页码: 9-15
ISSN:0034-6748
年代: 1969
DOI:10.1063/1.1683758
出版商: AIP
数据来源: AIP
摘要:
A new pulse amplitude discriminator, especially suitable for counting experiments, has been developed. Ten units were built and tested, and all of the threshold temperature coefficients are in the range −0.03 to +0.12mV per °C. At very high rates the threshold is stable and the output pulse width and the dead time change negligibly. Since the unit incorporates a number of convenient features, including provision for automatic dead‐time correction, it is an extremely versatile instrument.