首页   按字顺浏览 期刊浏览 卷期浏览 Resolution of time‐of‐flight mass spectrometers evaluated for secondary n...
Resolution of time‐of‐flight mass spectrometers evaluated for secondary neutral mass spectrometry

 

作者: Makoto Kato,   Akinori Mogami,   Motohiro Naito,   Shingo Ichimura,   Hazime Shimizu,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1988)
卷期: Volume 59, issue 9  

页码: 1947-1950

 

ISSN:0034-6748

 

年代: 1988

 

DOI:10.1063/1.1140056

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Mass resolution of a time‐of‐flight mass spectrometer with a two‐stage electrostatic reflector is calculated for secondary neutral mass spectrometry. The instrument parameters are optimized for energy and space focusing: correcting the flight time difference due to the energy width &Dgr;Eof sputtered particles and the spatial width &Dgr;sof an ionizing laser beam. The effect of &Dgr;scan be compensated by applying an acceleration field to the ionizing region, and the maximum resolution becomes about 1000 for &Dgr;E=10 eV and &Dgr;s=1.0 mm.

 

点击下载:  PDF (354KB)



返 回