Evaluation of mammography equipment performance, dose and image quality in five Latin American countries
作者:
M.-E. Brandan,
C. Ruiz-Trejo,
C. E. M. Caspani,
I. Fleitas,
R. de-la-Mora,
A. A. Miranda,
M.-C. Plazas,
C.-M. Betancourt,
C. Borras,
期刊:
AIP Conference Proceedings
(AIP Available online 1901)
卷期:
Volume 593,
issue 1
页码: 71-75
ISSN:0094-243X
年代: 1901
DOI:10.1063/1.1420468
出版商: AIP
数据来源: AIP
摘要:
Under the auspices of PAHO/WHO, a multicentric investigation is carried out in five Latin American countries. Its aim is to correlate quality indicators of radiology services with the accuracy of the radiological interpretation as determined by a panel of radiology experts. We present preliminary results from mammographic imaging facilities. Evaluation of the equipment performance and dose measurements in 21 mammographic units show that, on the average, 75&percent; of the units comply with recommendations issued by various organizations. An independent evaluation of the quality of the clinical images show strong variations among the different radiological services. ©2001 American Institute of Physics.
点击下载:
PDF
(559KB)
返 回