A dc strain‐gauge bridge amplifier configuration with automatic compensation of thermal drift using common operational amplifiers is presented. The amplifier operation is divided into two steps: one of auto‐zeroing and other of amplification of the input signal. The zeroing of the input signal is achieved by switching the bridge excitation to zero periodically at a frequency of about 100 Hz, thus limiting the maximum frequency of the input signal to about 8 Hz. With a gain of 5000, the output voltage drift of the amplifier is measured to be about 1 mV for a temperature change of 40 °C.

 