A Computer-Assisted Device for the Intraoperative CT-Correlated Localization of Brain Tumors
作者:
H. Reinhardt,
H. Meyer,
E. Amrein,
期刊:
European Surgical Research
(Karger Available online 1988)
卷期:
Volume 20,
issue 1
页码: 51-58
ISSN:0014-312X
年代: 1988
DOI:10.1159/000128741
出版商: S. Karger AG
关键词: Targeting device;Three-dimensional digitizer;CT-guided stereotaxy;Brain tumor;Radiophosphorus test;Semiconductor probe
数据来源: Karger
摘要:
In a series of 72 open brain tumor operations the 32P (radiophosphorus) test proved to be valuable for the removal of visually not detectable tumor residues. As during resection a topographical orientation in the depth of tumor cavities was nearly impossible, a computerized measuring device was developed which enables the surgeon to locate the tumor boundaries by simultaneous comparison with preoperative CT scans. For 32P-beta emission measurements a miniature semiconductor probe can be attached to the tip of the four-axis digitizing arm allowing the evaluation of radicality by topographic distinction between tumor ramifications and normal brain tissue. The device was tested during 3 craniotomies and the perspectives of computerized stereotaxy are discussed.