首页   按字顺浏览 期刊浏览 卷期浏览 Determination of Trace Copper in Cord Blood Serum from New Born Infants by Flameless AA...
Determination of Trace Copper in Cord Blood Serum from New Born Infants by Flameless AAS with Aerosol Deposition Technique

 

作者: Fu-Sheng Wei,   JoyT. Yang,   Fang Yin,  

 

期刊: Analytical Letters  (Taylor Available online 1983)
卷期: Volume 16, issue 7  

页码: 501-508

 

ISSN:0003-2719

 

年代: 1983

 

DOI:10.1080/00032718308065186

 

出版商: Taylor & Francis Group

 

关键词: Copper;Flameless AAS;Cord blood serum;Analysis

 

数据来源: Taylor

 

摘要:

This paper discusses the appropriate conditions for determining trace cooper by flameless atomic absorption spectrophotometry with the aerosol deposition technique. Seventeen samples of cord blood serum from the new born infants in the Hefei district of China were analyzed by this method. The range of trace copper in the samples is between 0.40-0.77 ppm, mean value 0.55 ppm. The recovery is over 90%. The relative standard deviation is 2.8%, determined by measurement of six times on a certain serum sample.

 

点击下载:  PDF (242KB)



返 回