首页   按字顺浏览 期刊浏览 卷期浏览 Countercurrent Chromatographic Separation of Proteins by Cross-Axis Coil Planet Centrif...
Countercurrent Chromatographic Separation of Proteins by Cross-Axis Coil Planet Centrifuge: Choice of Polymer Phase Systems and Revolution Speed

 

作者: K. Shinomiya,   Y. Kabasawa,   Y. Ito,  

 

期刊: Journal of Liquid Chromatography & Related Technologies  (Taylor Available online 1998)
卷期: Volume 21, issue 11  

页码: 1727-1736

 

ISSN:1082-6076

 

年代: 1998

 

DOI:10.1080/10826079808001255

 

出版商: Taylor & Francis Group

 

数据来源: Taylor

 

摘要:

Countercurrent chromatographic separation of proteins by the cross-axis coil planet centrifuge was maximized by selecting the suitable polymer phase system and revolution speed. Polymer phase systems composed of polyethylene glycol (PEG) 1000 and several inorganic salts were examined to determine partition coefficient values (K) for various proteins. The overall results indicated that the polymer phase system composed of 12.5% (w/w) PEG 1000 and 12.5% (w/w) dibasic potassium phosphate yielded suitable K values for most proteins except for cytochrome C and apo-transferrin which may be separable with a solvent system composed of 12.5% (w/w) PEG 1000 and 24% (w/w) potassium citrate. A series of experiments with the PEG 1000 - potassium phosphate system under various revolution speeds revealed that the best separation was achieved at 850 rpm. The above optimized conditions may be applied to separations of other protein samples.

 