Determination of phenolic pollutants in water using permeation sampling
作者:
Guo‐Zheng Zhang,
JamesK. Hardy,
期刊:
Journal of Environmental Science and Health . Part A: Environmental Science and Engineering
(Taylor Available online 1989)
卷期:
Volume 24,
issue 3
页码: 279-295
ISSN:0360-1226
年代: 1989
DOI:10.1080/10934528909375481
出版商: Taylor & Francis Group
关键词: Phenolic Pollutants;Permeation;Membrane;Adsorbents
数据来源: Taylor
摘要:
A passive sampling technique for the determination of time‐weighted‐average concentrations of phenolic priority pollutants in water is described. Phenols permeate a silicone polycarbonate membrane and are collected on a polymeric adsorbent. For high concentration ranges, phenols are adsorbed on the XAD‐7 resin and then desorbed with acetonitrile. Low ppb concentrations of phenols are analyzed by collection on Tenax‐TA followed by thermal desorption. Separation and quantification is achieved by capillary column gas chromatography using a flame ionization detector. A linear relationship between the amount of phenols collected and the products of the exposure time and phenolic concentrations over the range of 5 ppb ‐ 20 ppm have been determined. Temperature and solution pH affect permeation rates of the phenols. The sampling method is simple, inexpensive and suitable for multilocation monitoring.
点击下载:
PDF (471KB)