首页   按字顺浏览 期刊浏览 卷期浏览 Comparison of a Multipore Column with a Mixed-Bed Column for Size Exclusion Chromatogra...
Comparison of a Multipore Column with a Mixed-Bed Column for Size Exclusion Chromatography

 

作者: M. Nagata,   T. Kato,   H. Furutani,  

 

期刊: Journal of Liquid Chromatography & Related Technologies  (Taylor Available online 1998)
卷期: Volume 21, issue 10  

页码: 1471-1484

 

ISSN:1082-6076

 

年代: 1998

 

DOI:10.1080/10826079808000528

 

出版商: Taylor & Francis Group

 

数据来源: Taylor

 

摘要:

A new type of column packed with polystyrene gels which have a broad pore size distribution in a particle (henceforth referred to as the multipore column) was developed, and its performance was compared with the so-called linear column (henceforth referred to as the mixed-bed column). The chromatogram of the epoxy resin (Epikote 1009) on the multipore column showed a monomodal distribution, but the chromatogram on the mixed-bed column showed the inflection on the chromatogram at the molecular mass of about 10,000 as polystyrene equivalent. There was a big difference in the pore size distributions of gels packed in the two types of columns; the pore size distribution of the multipore column showed the trapezoid, while that of the mixed-bed column had sharp maxima for pores with a diameter of 0.08 μm. This difference in the pore size distribution may be the main reason for the inflection phenomenon, though the calibration curves of these two types of columns were almost the same. The multipore column also had the advantage of the mixed-bed column in the linearity of the calibration curve.

 