Molecular Weight-Retention Volume Relation for Narrow and Broad Polystyrenes by Size Exclusion Chromatography Equipped With a Light Scattering Detector
作者:
S. Mori,
M. Ishikawa,
期刊:
Journal of Liquid Chromatography & Related Technologies
(Taylor Available online 1998)
卷期:
Volume 21,
issue 8
页码: 1107-1117
ISSN:1082-6076
年代: 1998
DOI:10.1080/10826079808006587
出版商: Taylor & Francis Group
数据来源: Taylor
摘要:
The difference of the relation of molecular weight (MW) and retention volume of narrow polystyrene (PS) samples from broad ones in the size exclusion chromatography - light scattering (LS)/refractive index (RI) detectors is discussed. Effects of interdetector delay volume on the relation was serious and the approach to adjust the interdetector delay volume so as to get a plateau relation of MW and retention volume for PS samples with a narrow MW distribution was proposed. The interdetector delay volume was found to be a function of MW. PS samples with a broad MW distribution had the similar relation of MW and retention volume to the calibration curve constructed with PS standards with narrow MW distributions; and the MW averages obtained by the LS/RI approach gave the comparable values to those calculated by a calibration curve method.