Adaptation of an Automated Assay for Determination of β‐Hydroxybutyrate in Dogs Using a Random Access Analyzer
作者:
Mary M. Christopher,
Jacqueline L. Periera,
Robin L. Brigmon,
期刊:
Veterinary Clinical Pathology
(WILEY Available online 1992)
卷期:
Volume 21,
issue 1
页码: 3-8
ISSN:0275-6382
年代: 1992
DOI:10.1111/j.1939-165X.1992.tb00573.x
出版商: Blackwell Publishing Ltd
关键词: ketone bodies;ketones;diabetes mellitus;clinical chemistry
数据来源: WILEY
摘要:
SummaryAn automated method for measuring β‐hydroxybutyrate was adapted to the Ciba‐Corning 550 Express™ random access analyzer. The assay was based on a kinetic reaction utilizing hydroxybutyrate‐dehydrogenase. β‐hydroxybutyrate concentration (mmol/L) was calculated ratiometrically using a 1.0 mmol/L standard. Canine serum, plasma, and urine were used without prior deproteinization and only a 30‐μl sample was required. The method demonstrated good linearity between 0 to 2 mmol/L of β‐hydroxybutyrate. Analytical recovery (accuracy) within these concentrations ranged from 85.8 to 113.3%. Both within‐run and day‐to‐day precision were determined, as was specificity of the assay in the presence of a variety of interfering substances. The automated assay was rapid and economical, with reagent stability maintained for at least 2 weeks at 4° C. This assay can readily be applied toward the assessment of ketoacidosis in dogs, and with further v