An Electrochemical Flow Cell For The Detection Of Metal Complexes
作者:
StevenA. Cohen,
DuaneE. Bartak,
期刊:
Analytical Letters
(Taylor Available online 1983)
卷期:
Volume 16,
issue 6
页码: 429-441
ISSN:0003-2719
年代: 1983
DOI:10.1080/00032718308065179
出版商: Taylor & Francis Group
关键词: Flow Cell;Reticulated Vitreous Carbon Electrode;Metal Complex Separation;Modified Gel Permeation Chromatography
数据来源: Taylor
摘要:
An electrochemical flow cell has been developed using reticulated vitreous carbon (RVC) coated with a thin film of mercury. The flow cell and electrode components are inexpensive and easily constructed. Reproducible results have been obtained for concentrations as low as 100 ppb. Peak currents were dependent on flow rate, electrode size, and mercury deposition time. Oxygen interference was reduced by nitrogen purging to enable analysis of sub-ppm metal concentrations. The cell was tested as a detector for metal complexes eluated from a modified gel permeation chromatographic column.