A DC-10 MHz Low Noise Amplifier System for Fast Photoconduction Decay Measurements under the Influence of Strong Noises Arising from a Light Pulse Generator
作者:
Jin Mizuguchi,
期刊:
Instrumentation Science & Technology
(Taylor Available online 1979)
卷期:
Volume 9,
issue 3
页码: 219-225
ISSN:1073-9149
年代: 1979
DOI:10.1080/10739147908543468
出版商: Taylor & Francis Group
数据来源: Taylor
摘要:
A pulse pre-amplifier for use in the measurements of fast photoconduction decays less than 1 us has been constructed by using ultrafast operational amplifiers. A flat frequency response with little phase distortion is obtained over the range from DC to 10 MHz and the rise time is about 30 ns. The amplifier should be operated with batteries in order to reduce the disturbance from strong noises of a light pulse generator used for the sample illumination.